The following illustrates a standard operating procedure (SOP) for applying a heel skive in FotoView.
Prerequisites
3D foot scan aligned in FotoView
Refer to the following guide for more information:
Goals
- Apply a Heel Skive in FotoView
Procedure
Click Adjust Mesh → Any View
Click Close from the Adjust Mesh toolbar
The Skive menu option will now be selectableClick Skive from the top menu
A Skive Plate will now appear on the display in an orientation specified by the Depth and Degree values specified on the Machined EVA prescription.The Skive Plate controls will also appear, which provides fine control over the position of the Skive Plate
Use the Skive Plate controls to orientate the Skive Plate according to the prescriptive variables
Skives can be requested on the Machined EVA prescription form using the Depth/Degree Method and also the STJ Axis Method. Please reference the individual guides below for assistance.
Depth/Degree Method
When Depth/Degree Method is selected on the Machined EVA prescription form, it will be visible under the 3D Scan Alignment header, within the Skive row.
Follow Steps 1-4 listed above
Click View → Back
Check the Degree of your Skive Plate using the Measure → Angle tool
Click View → Right Side (left foot) / View → Left Side (right foot)
Use the Pitch Fore/Aft control to match the specified Inclination
Orientate the Skive Plate so that you can accurately measure the Depth
Check the Depth of your Skive Plate using the Measure → Distance tool
Your 3D scan should now be aligned in FotoView, and you should be ready to generate your orthosis model.