The following illustrates a standard operating procedure (SOP) for aligning a 3D foot scan in RapidShell3DFotoView.

...

Table of Contents

Prerequisites

  • Prescription loaded into RapidShell3D FotoView

Refer to the following guide for more information:

📋 SOP: Loading Prescription in RapidShell3DFotoView

Goals

  •  Align a 3D foot scan in RapidShell3D FotoView

Procedure

  1. Click Files → Display Left Foot

    Image RemovedImage Added

  2. A 3D scan of the patient’s left foot should now appear on the screen

    Image Removed

    Click Place Points

    Image RemovedImage Added

  3. Click Alignment → Align Positive (Envisic)

    (info) If you are not capturing scans with the LaserCAM VPS 3D Scanner, then you may need to select one of the alternative alignment options. Your specific 3D scanner should capture foot scans with a standard rotation and translation relative to the global axes within FotoView, so discover the optimal alignment method through trial and error.

    Image Added

  4. In specific order, click the Heel Point → 1st MPJ Centre Point → 5th MPJ Centre Point

    Image RemovedImage Added

  5. Click Align for Valgus or Align for Varus

    (info) The prescription form may contain values for Forefoot Varus or Forefoot Valgus. If such values exist, please click the corresponding button.

  6. Click Accept

  7. Click Apply

    (info) The Distances box will display a distance between the Heel Centre point and 1st MPJ centre point that were placed on the 3D scan surface during Step 4. If a prescription value exists for Heel Centre to 1st MPJ (mm)a Scale Factor will appear and allow for automatic scaling of the 3D scan. Please note that it is not recommended to scale 3D scans captured with the LaserCAM VPS 3D scanner, or any other scanner that is known to capture highly accurate 3D foot scans.

    Position

  8. the scan in the Plantar view using the Use the Adjust Mesh controls along the bottom toolbar

    Image Removed

    Image Removed

    Position the scan in the Heel view to translate the scan so that is aligned according to the yellow foot profile.

    Image Added

    Image Added

  9. Click Adjust Mesh → Heel and position the scan using controls along the bottom toolbar according to a provided Acceptable Reference Measurement.

    Method 1: Heel Bisection Measurement

    📋 SOP: Heel Bisection Measurement

    You may use the blue Z-axis line to reference the heel bisection in neutral (0°). You can then use the Invert and Evert controls from the bottom toolbar to rotate the scan into the alignment specified in the Scan Angle field on the prescription.

    Image RemovedImage Added


    Method 2: Forefoot Varus/Valgus Alignment Method

    📋 SOP: Forefoot Varus/Valgus Measurement

    If a Forefoot Varus or Forefoot Valgus measurement has been recorded on the prescription form, RapidShell3D FotoView will automatically rotate the 3D scan to this value. You will need to measure the alignment to ensure that you have accurately rotated the 3D scan.

    Image RemovedImage Added

    Image RemovedPosition the scan in the Medial view Image Added

  10. Click Adjust Mesh → Side and position the scan so that the blue Z-axis line is breaking through the heel of the scan at approximately 20mm in height

    Image Removed

    30mm in height.

    (info) The yellow profile bars that extend vertically are 30mm in height, which should be used as a reference.

    Image Added

Tip

Your 3D scan should now be aligned in RapidShell3D, and you should be ready to generate your orthosis model.FotoView